Measuring Samples in a 96 Well Plate
Users can measure all of the samples in a 96 well plate in about 6 minutes using the Thermo Scientific NanoDropTM 8000 Spectrophotometer.
View demo video (.wmv format)
Sample Pedestal Reconditioning
The sample pedestals occasionally become "unconditoned" causing the sample liquid column not to form properly. The software will alert the user when this occurs and a "reconditioning" procedure will be required (see images below). A reconditioning kit (PR-1) can be purchased from NanoDrop or your local distributor for reconditioning the sample pedestal surfaces.
View reconditioning demonstration video (.wmv format)
| |
|
 |
|
|
| |
Image of unconditioned sample pedestals- sample spreads out on surface |
|
Image of reconditioned sample pedestals- sample "beads" up on surface |
|
|