Instrument Type
|
Spectrophotometer |
|
|
Minimum Sample Size |
0.5 µl |
|
Sample Number |
1 |
|
Path Length |
1 mm (auto-ranging to 0.05 mm) |
|
Light Source(s) |
Xenon flash lamp |
|
Detector Type |
2048 - element linear silicon CCD array |
|
Wavelength Range |
190 - 840 nm |
|
Wavelength Accuracy |
1 nm |
|
Spectral Resolution |
≤ 1.8 nm (FWHM at Hg 253.7 nm) |
|
Absorbance Precision |
0.002 (1 mm path) |
|
Absorbance Accuracy |
3% (at 0.74 at 350 nm) |
|
Absorbance Range |
0.04 - 300 (10 mm equivalent) |
|
Detection Limit |
2 ng/µl (dsDNA) |
|
Maximum Concentration |
15,000 ng/µl (dsDNA) |
|
Measurement Time |
< 5 seconds |
|
Footprint |
14 x 20 cm |
|
Weight |
2.1 kg |
|
Sample Pedestal Material of Construction |
303 stainless steel and quartz fiber |
|
Operating Voltage |
12 vdc |
|
Operating Power Consumption |
12 - 18 W (max 30 W) |
|
Standby Power Consumption |
5 W |
|
Software Compatibility |
Professional versions of Windows 7 (32 bit and 64 bit), Windows Vista (32 bit), Windows XP (32 bit) |
| |
|
|
Cuvette specifications |
|
| |
|
|
Z-Height |
8.5 mm |
|
Heating |
37 ± 0.5 °C |
|
Stirrer: |
150-850 rpm |
|
Path Length |
10, 5, 2, 1 mm |
|
Absorbance Range |
0.008 - 1.5 |
|
Detection Limit |
0.4 ng/µl (dsDNA) |
|
Maximum Concentration |
750 ng/µl (dsDNA) |
|
Measurement Time |
< 3 seconds |
| Cuvette Dimensions |
12.5 mm x 12.5 mm, up to 48 mm H |
| Type |
Masked cuvette |
| UL/CSA and CE |
All units are approved to these standards |